Start-up:
The DF-7010 performs a self-test once the operating voltage as been con-
nected. Ia = 3 mA [adjustable];
After 5 minutes the device switches to
measuring mode.
In the Ex-area: Before the terminal section is opened, the measuring head
must be isolated from the voltage supply.
Status signals:
Calibration
3 mA; fault
2 mA: Cable break
0 mA [standard]
Function test:
Test gas function via: Integrated test gas nozzles at 0.5 l/min [for standard
test gas] or test cap at 1.0 l/ min, [Part No. D6079762] or splash water hous-
ing SG 70 at 1 l/min or pump adapter PA 70 at 1.0 l/min
Calibration:
The unit is freshly calibrated or re-calibrated onsite using the
AUER CONTROL module. Operating Instruction AUER-CONTROL
Part No.: D0792125.
Approved measuring components, calibration conditions, measuring ranges,
alarm points and linearization tables according to materials list: D0792421.
Other measuring components available on request.