MSA AUER
MSA
Appendix
Control Unit Models 9010 LCD and 9020 LCD
65
GB
9010/20 LCD
Sensor data sheet DF-7010
Part No.: D0792602
Terminal Block
9010/9020
Sensor Type DF-7010
Caution:
Before connecting the measuring head, check hardware and software
settings.
Sensor type:
Catalytic/active/3-wire/4–20 mA linear/current source [3 wire active]
Connection data:
Supply voltage:
16 – 32 V DC
Maximum current draw:
120 mA
Power requirement:
4.0 W [excluding cable length]
Connection cable:
3-lead, 80% shielded
Maximum resistance:
200 Ohm
Maximum cable length:
1,000 m [with 1.5 mm
2
lead cross-section]
Cable diameter:
8 – 13 mm
Permitted lead cross-section:
0.5 – 2.5 mm
2
Cable entry:
M20 x 1.5
Conditions for
use:
Assembly:
Wall-mounted
Protection:
IP 42 to EN 60529
Explosion protection:
II 2G EEx d m e ib IIC T6
Approvals:
DMT 99 ATEX E 023 X
Temperature:
-20 °C to +40 °C
Humidity:
5 – 95 % rel. humidity; non-condensing
Pressure:
950 – 1,100 hPa
Weight:
approx. 1.5 kg
Dimensions W x D x H:
150 mm x 88 mm x 158 mm
For further details see the Operating and Maintenance Instructions.