USER MANUAL AMPRO
plus
MRU GmbH, D-74172 Neckarsulm
11 / 100
MRU GmbH
Fuchshalde 8 + 12
74172 Neckarsulm - Obereisesheim
GERMANY
Fon +49 71 32 99 62 0 (Zentrale)
Fon +49 71 32 99 62 61 (Kundendienst)
Fax +49 71 32 99 62 20
Email:
Internet:
www.mru.eu
2.3.
Important general information (EN 50379 and VDI 4206)
The analyzer is not designed to be used for continuous measurements
Before using the analyser verify the condition of the various parts of the
analyser, such as the probe, the ambient air conditions, the condensate
separator, star filter and the connectors for damage and/or blockages.
When starting the analyser up it will take between 1 – 3 minutes to set to
zero depending on the condition of the sensors and ambient conditions.
(Zeroing).
The minimum zeroing time of the analyser to achieve correct measure-
ment values can be expected by 1.5 minutes!
Exposure to acids; aggressive gases such as Sulphur; vapours such as thin-
ners, gasoline, alcohol and paint, etc. can damage, reduce the life of, or
destroy the sensors
Die Lebenserwartung der Sensoren beträgt, abhängig von Nutzung,
Wartung und Pflege des Messgerätes bei dem O
2
Sensor ca. 2 Jahre und
bei dem CO Sensor ca. 2-3 Jahre.
The life of the sensors depends on how they are used, maintained and
treated. Typical average life expectations are: O2 - 2 years; CO - 2 - 3 years
and NO - 3 years.
Die Lebensdauer des Akkus beträgt mindestens 500 Lade- Entladezyklen.
Mit zunehmender Zyklenzahl reduziert sich die in den technischen Daten
angegebene Gerätebetriebsdauer pro Akkuladung.
The use of the analyzer for regulatory purposes is subject to special regu-
lations (for example a periodical examination of the analyzer). Please ob-
tain the appropriate regulations from your local responsible authority.
2.4.
Packaging
Save the original carton and packing materials to prevent damage in transit
in case you need to return the unit to the factory.
2.5.
Return of hazardous materials
Waste Disposal/Returns/Warranty - MRU is required to accept the return of
hazardous waste such as electro-chemical sensors that cannot be disposed
of locally.
Hazardous waste must be returned to MRU prepaid.