Performing measurement
In the basic configuration, each analyser has the complete functionality you
need for gas measurement.
The process of gas measurement is described below.
The description of other optionally available measuring programs can be
found in the appendix or on separate supplementary sheets.
7.1.
Perform Gas measurement
Core flow search
You can preselect whether a core flow search is to be carried out in the
measuring programs before the start of the measurements.
The core flow search will help you to find the optimal measurement point in
the stack. The core flow can be identified by the maximum flue gas tempera-
ture. In high reaction time the analyser displays the trend of the flue gas
temperature. Insert the probe pipe slowly into the stack and position your
probe tube when you have reached the maximum flue gas temperature that
is displayed
Positioning the probe in the core flow:
NOTE
Note that you have to switch on the core flow search in the
menu Measurement settings before starting a gas meas-
urement.
See also chapter 5.4 Set measurement, Page 26