Service setup
Page 53
9.1.8
First start
-
Press the button
„FirstStart”
-
This is a tool where the system can perform the
initialization process manually (see chapter 4.11)
without restarting the system
9.1.9
Direct access
-
Press the button
„Direct Access”
-
This is a tool where each components functionality
can be checked manually
o
Actuators:
P1: pump 1 (input water)
P2: pump 2 (pure water)
12/24V: voltage of P2
(EDI): electro-deionized cell (only
Aquinity²E
)
UV PW: UV-lamp in pure water
SV01~SV06
: solenoid ventile 1…6
o
Digital input:
PS1: pressure sensor of input water
Regen.: self-regeneration EDI-cell
Swimmer: overflow detector
Tap (NC): dispenser (normal-close)
T.High: tanklevel sensor (top) (only
Aq²P10
)
T.Low: tanklevel sensor (bottom) (only
Aq²P10
)
Salt: salt detector
Softener: softener detector
o
Analog input:
Q1-Q4: conductivity sensor
1…4
T2-T4: temperature sensor
2…4
Tank