System Features
membra
Pure
GmbH
Wolfgang-Küntscher-Str.14
16761 Hennigsdorf
Germany
Tel.: +49 (0) 3302/201-20 0
Fax.: +49 (0) 03302/201-20 21
E-Mail: [email protected]
Internet: www.membrapure.de
page 9
3.2
Series overview
Aquinity² E
model
UV-
reactor
UF-
module
TOC-
monitoring
Art.-No.:
(35L-tank)
Art.-No.:
(70L-tank)
Reagent
○
○
○
114-0053
114-0063
Analytical
●
○
○
114-0054
114-0064
LifeScience
●
●
○
114-0055
114-0065
Analytical TI
●
○
●
114-0058
114-0068
LifeScience TI
●
●
●
114-0059
114-0069
Aquinity²
P
model
UV-
reactor
UF-
module
TOC-
monitoring
Art.-No.:
(35L-tank)
Art.-No.:
(70L-tank)
Reagent
○
○
○
114-0050
114-0060
Analytical
●
○
○
114-0051
114-0061
LifeScience
●
●
○
114-0052
114-0062
Analytical TI
●
○
●
114-0056
114-0066
LifeScience TI
●
●
●
114-0057
114-0067
Table 1: Series-overview
3.3
Technical data
Aquinity² E
RE
AL
LS
AL TI
LS TI
Product water specifications
product water flow
L/min
up to 2
conductivity at 25°C
µS/cm
0,055
resistivity at 25°C
M
Ω
cm
18,2
TOC-concentration
ppb
< 10
< 1
< 5
< 1
< 5
Endotoxin-concentration
EU/ml
n/a
n/a
< 0,001
n/a
<0,001
System specifications
shipping weight (35L-tank)
kg
shipping weight (70L-tank)
kg
power supply
V
100 … 240
Hz
50 / 60
dimension (W/D/H)
cm
35 / 54 / 51
dimdispenser
(W/D/H)
cm
35 / 75 / 63
Table 2: Technical Data