Service setup
Page 61
9.5
Software parameters
9.5.1
Configuration file (config.mcf)
Ini-key
Default value
Help
b_all_tools
F
TRUE -> Change all Packs is active
b_aquinity_start
F
TRUE -> Water production is running
b_lf_sensor_1
F
TRUE -> LF1 is in working range
b_lf_sensor_2
F
TRUE -> LF2 is in working range
b_load_save_backup
F
Not used anymore
b_tank_level_limit
F
TRUE -> Tank is full
b_temp_too_high
F
TRUE -> T3 is over limit
dbl_init_q1_value
50
Init value of Q1 Sensor
dbl_init_q2_value
5
Init value of Q2 Sensor
dbl_init_q3_value
0,055
Init value of Q3 Sensor
dbl_init_q4_value
0,055
Init value of Q4 Sensor
dbl_init_tank_level
25
Init value of Tank Sensor
dbl_init_toc_value
3
Init value of TOC (abs (Q3-Q4))
dbl_q_tank_volt_limit_warning
0,01
(0,01V) on all sensors -> COM failure!
dbl_q1_set_limit_min
1
Q1 < 1µS -> Error
dbl_q2_set_limit_min
0,035
Q2 < 0.035µS -> Error
dbl_q3_set_limit_min
0,035
Q3 < 0.035µS -> Error
dbl_sec_back_to_main
1200
After 2 mins back to main
dbl_sec_call_memory
60
Refresh memory display (1 min)
dbl_sec_check_op_hours
600
Update operating hours (10 mins)
dbl_sec_check_p1_counter
1800
Check P1 counter (30 min)
dbl_sec_count_op_hours
60
Counter for operating hours (1 min)
dbl_sec_data_high_density
3600
Acquire data every 1 hour
dbl_sec_data_low_density
86400
Acquire data every 24 hours
dbl_sec_data_mid_density
43200
Acquire data every 12 hours
dbl_sec_flush_change_cm
3600
Init flushing Con.M. (1 hr)(AQ²E only)
dbl_sec_flush_change_cm_edi
3600
Flushing Con.M. after change (1 hr)
dbl_sec_flush_change_pp
300
Init flushing PropPak (5 min)
dbl_sec_flush_cm_go_idle
0
Flush Con.M. when tank full (0 sec)
dbl_sec_flush_edi_go_idle
300
5 mins flush EDI when Tank full
dbl_sec_flush_ro_go_idle
60
Flush RO when tank full (1 min)
dbl_sec_flush_upw_uf
300
5 mins flush UF after change
dbl_sec_flush_upw_uf_sv6_close
15
Flushing PW -> SV6 close (15 sec)
dbl_sec_flush_upw_uf_sv6_open
60
Flushing PW -> SV6 open (15 sec)
dbl_sec_init_flush_upw_no_uf
300
5 mins first flush UPW-Pack
dbl_sec_init_flush_upw_uf
2700
45 mins first flush UF module
dbl_sec_init_uf_sv6_close
10
Init flushing PW -> SV6 close (10 sec)
dbl_sec_init_uf_sv6_open
10
Init flushing PW -> SV6 open (10 sec)
dbl_sec_io_command_hs
30
Handshake with control PCB (30 sec)
dbl_sec_message_switcher
2
Switch Messages every (2 sec)
dbl_sec_no_comm_warning
120
Time to show error NO COMM (2 mins)
dbl_sec_no_ps1_over
10
PS1 problem is over, 10s wait, back
dbl_sec_no_ps1_sv1_close
5
PS1 problem, close SV1 for 5s
dbl_sec_no_ps1_sv1_open
2
PS1 problem, open SV1 for 2s
dbl_sec_p2_12_active
300
Recirculating time (5 min) (when full)
dbl_sec_p2_12_sleep
10800
System sleep time (3 hr) (when full)
dbl_sec_ps1_error
1800
30 mins Pressure problem -> STOP
dbl_sec_ps1_warning
15
15 secs Pressure problem -> Warning