System Control Panel
membra
Pure
GmbH
Wolfgang-Küntscher-Str.14
16761 Hennigsdorf
Germany
Tel.: +49 (0) 3302/201-20 0
Fax.: +49 (0) 03302/201-20 21
E-Mail: [email protected]
Internet: www.membrapure.de
page 32
5.3.3
Logfile
-
Press the button
„
LOG
“
-
The logfile will be displayed
5.4
User setup
In the user setup values such as limitations for the conductivity sensors, tank sensor or
system time setting can be adjusted.
-
To enter the user setup please enter the password
9999
and confirm entry with “OK”
5.4.1
Limits setup
-
Press the button
„Limits”
-
Change the value of “Stop Production Tank” to adjust
upper tank level (stop production)
-
Change the value of
“Start Production Tank”to adjust
the lower tank level (switch on production)
-
Set Q3-limit: max. value of conductivity Q3 without
showing any warning
-
Conductivity can be displayed in two different ways
(µS/cm or MΩ*cm)
-
Press.Switch ON/OFF is to activate the monitoring of
pressure sensor switch from feed water
-
Interval ON/OFF is to define how the system works
when the tank is already full
Interval ON: system recirculates every 3 hours
Interval OFF: system recirculates permanently