Care and maintenance
membra
Pure
GmbH
Wolfgang-Küntscher-Str.14
16761 Hennigsdorf
Germany
Tel.: +49 (0) 3302/201-20 0
Fax.: +49 (0) 03302/201-20 21
E-Mail: [email protected]
Internet: www.membrapure.de
page 43
Figure 21: Disinfection cell with mounted strainer
-
Attach the strainer with the ClO
2
-tablet into the housing of the disinfection cell. To
do that press the strainer into the fit inside of the housing.
-
Close the disinfection cell by turning the cell case onto the cell-housing
counterclockwise
6.8
Replacement of reverse osmosis module
-
Turn off the system by unplugging the power adapter out of the electrical outlet.
-
Disconnect the system from feed water.
-
Open the system.
-
Remove reverse osmosis housing from its holder.
-
Remove tubes from the reverse osmosis housing (caution: water leakage).
-
Open the housing and remove the reverse osmosis module.
-
Unpack the new reverse osmosis module and put it into the housing.
-
Close the housing.
-
Reconnect the tubes onto the reverse osmosis housing.
-
Plug the reverse osmosis housing back into its holders.
-
Connect the system to feed water.
-
Turn on the system by plugging the power adapter into an electrical outlet.
-
Flush the reverse osmosis module
by start the tool “
CHANGE ProPak
.”
(see chapter
5.2.1).
-
Check impermeability on all connections
-
Close the front door, the system is ready for operation.