Installation and start up
membra
Pure
GmbH
Wolfgang-Küntscher-Str.14
16761 Hennigsdorf
Germany
Tel.: +49 (0) 3302/201-20 0
Fax.: +49 (0) 03302/201-20 21
E-Mail: [email protected]
Internet: www.membrapure.de
page 20
Figure 11: Sealing plug
4.6
Connections
All waste tubes must be led with a downward gradient toward the waste water drain. Any
reduction of the diameter, loops, siphons or breaks in the line must be avoided.
A minimum safety distance of 2 cm (3/4-inch) must be between tube end and waste water
to prevent siphoning. An air gap of a minimum distance of double inner tube diameter is
recommended.
-
Plug the tubes in until stop position. The tube must bottom out in connecter. Tubes
have to be locked with the union nut.
-
Connect the tubes correspondently to their markings and push them to the stop.
-
Plug the socket of the level sensor to the tank and fasten it.
-
Screw the tap to the provided opening at the front of the tank. Please note that the
tap is supposed to point downwards.
-
Screw the cap onto the tank
2. Depress fastening ring
3. Remove sealing plug
1. remove union nut