www.miinet.com
Moore Industries-International, Inc.
- 96 -
User’s Manual
238-741-00K
May 2019
Programmable Smart HART Temperature Transmitter
THZ
3
/ TDZ
3
SECTION 5
Display Menu (TDZ
3
only)
The Display menu allows you to select which variables are shown and also allows for custom
names and engineering units (EGU). Refer to Figure 5.29 for further clarification.
Figure 5.29.
TDZ
3
Display Menu Parameters Analog Output
When PV, SV, TV, or QV are selected for display, each will result in three screens.
123.45 / Dynamic Var Name
123.45 / Source Device Var Name
123.45 / Units
When the Percent of Range or Loop Current are selected for display, each will be shown
using two screens:
12.345 / AO
12.345 / MA
or
12.345 / PCT
12.345 / PCT
When selected, a Device Variable is shown using two screens.
123.45 / Name (default or custom name)
123.45 / UNITS
If there is a problem with the underlying Device Variable, only one screen will be shown:
BADDQ / Dynamic Var Name
In this last case, the nature of the error will be shown separately.