www.miinet.com
Moore Industries-International, Inc.
- 23 -
User’s Manual
238-741-00K
May 2019
Programmable Smart HART Temperature Transmitter
THZ
3
/ TDZ
3
Figure 2.11.
Output Trimming
The output of the THZ
3
/TDZ
3
is accurately calibrated prior to delivery to our customers.
However, you may wish to trim the instrument’s analog output to match what you are reading
on your monitoring or control system. By example: if the THZ
3
is putting out 12.0mA and your
DCS is reading 11.85mA you can trim the output of the THZ
3
so that the two readings will match.
Note: trimming of the analog output will not affect the HART digital data.
To perform output trimming set up your THZ
3
/TDZ
3
as shown in Figure 2.11.
12-24Vdc
Power
Supply
+
-
ADDR
ADDR
A
TDZ
3
Voltmeter
Load= 250 ohm
Moore Industries
P/N: 804-030-26
To USB
Port of PC
PC
HART Modem
connects to
Serial/USB
COM Port of PC
You can then perform Sensor 2 Trim if you are using two input sensors.
Your trim values will be automatically saved to your instrument.
SECTION 2