Chapter 9: Maintenance and Troubleshooting
5032022 C
137
3. Press down on the black top of the pinch valve and then slide the dispense tubing out of
the slot.
Figure 9-11: Dispense Tubing in Pinch Valve
4. Disconnect the blue quick-fit connector on the dispense tubing from the blue connector
on the top of the instrument.
5. Drain the dispense tubing into an appropriate receptacle, and then set the dispense
tubing aside.
6. Remove the aspiration tubing from the red connector on the manifold.
7. Disconnect the red quick-fit connector on the aspiration tubing from the red connector
on the top of the instrument.
8. Drain the aspiration tubing into an appropriate receptacle, and then set the aspiration
tubing aside.
9. Lift the manifold off the horizontal manifold holder
10. Drain the manifold into an appropriate receptacle, and then set the manifold aside.
11. Close the clear cover.