Chapter 9: Maintenance and Troubleshooting
5032022 C
131
To replace the fuses:
WARNING! HIGH VOLTAGE
Power off the instrument and disconnect the
power cord before you do maintenance procedures that require removal of a
panel or cover or disassembly of an interior instrument component.
1. Switch the power switch on the rear of the instrument to the off position.
2. Unplug the power cord from the power port.
3. Place a small slot-head screwdriver below the fuse carrier and gently pry the fuse carrier
from the instrument.
4. Gently pull the old fuses from the carrier by hand.
Figure 9-6: Fuse Carrier
5. Gently place new fuses into the carrier by hand.
6. Return the fuse carrier to the slot above the power port with the tabs on the carrier
pointing down.
7. Press the fuse carrier into the instrument until the carrier snaps into place.
8. Plug the power cord into the power port.
9. Turn on power to the instrument.
Note:
If the instrument still does not power on after changing the fuses, contact
Molecular Devices technical support. See
.