Mi Microplate Washer User Guide
28
5032022 C
Confirm the Level of the Waste Bottle
If the
Waste Warning
option is enabled in the
Setup
menu, then the software displays a
prompt to make sure that the waste bottle has sufficient room for aspirated liquid during the
selected procedure.
CAUTION!
Running a procedure without sufficient room in the waste bottle can
damage the instrument.
To enable the
Waste Warning
option in the
Setup
menu, see
.
Figure 3-3: The Waste Empty prompt
Button
Action
On the left
If the waste bottle is not empty, press this button to return to
Run Procedure
. See
Running a Procedure on page 25
.
On the right
If the waste bottle is empty, press this button to go to the next step. If no other
options have been enabled, then the selected procedure starts to run. See
Monitoring the Procedure on page 30
The button has no function for this option.
The button has no function for this option.
The button has no function for this option.
Table 3-3: Actions of Control Panel Buttons
Summary of Contents for MultiWash+
Page 1: ...5032022 C August 2019 MultiWash Microplate Washer User Guide ...
Page 20: ...MicroWash Microplate Washer User Guide 20 5032022 C ...
Page 50: ...MicroWash Microplate Washer User Guide 50 5032022 C ...
Page 64: ...MicroWash Microplate Washer User Guide 64 5032022 C ...
Page 144: ...MicroWash Microplate Washer User Guide 144 5032022 C ...