Chapter 7: Defining a Microplate Type
5032022 C
109
Define the Bottom Dispense Height
The bottom dispense height defines the height of the manifold during a dispense for a
Bottom
wash. The dispense height must be higher or equal to the height for aspiration.
Figure 7-41: Bottom dispense height
You can define the bottom dispense height by editing the offset value, or by using the
"teach-in" method.
Figure 7-42: The Bottom Dispense Height prompt
Button
Action
On the left
Press this button to end the process and return to
Define Plate
without saving any
changes. See
Defining a Microplate Type on page 77
.
On the right
Press this button to move the microplate to the currently saved bottom dispense
position and switch to "teach-in" mode. See
Teach-In the Bottom Dispense Height on
Press this button to increase the offset value by 0.1 mm to a maximum of 18.8 mm.
Press this button to decrease the offset value by 0.1 mm to a minimum of 0.0 mm.
Press this button to accept the displayed offset value and then define the overflow
dispense height. See
Define the Overflow Dispense Height on page 111
.
Table 7-31: Actions of Control Panel Buttons
Summary of Contents for MultiWash+
Page 1: ...5032022 C August 2019 MultiWash Microplate Washer User Guide ...
Page 20: ...MicroWash Microplate Washer User Guide 20 5032022 C ...
Page 50: ...MicroWash Microplate Washer User Guide 50 5032022 C ...
Page 64: ...MicroWash Microplate Washer User Guide 64 5032022 C ...
Page 144: ...MicroWash Microplate Washer User Guide 144 5032022 C ...