Chapter 7: Defining a Microplate Type
5032022 C
103
Define the Second Side for Crosswise Aspiration
This option is available only for flat-bottom microplates with crosswise aspiration enabled.
This option defines the position of the microplate for the second step of a crosswise
aspiration. Molecular Devices recommends that you place the aspiration needle close to the
edge that is opposite from the first aspiration position.
Figure 7-32: The second side for crosswise aspiration
You can define the second side for crosswise aspiration by editing the offset value, or by
using the "teach-in" method.
Figure 7-33: The Side Position Rear prompt
Button
Action
On the left
Press this button to end the process and return to
Define Plate
without saving any
changes. See
Defining a Microplate Type on page 77
.
On the right
Press this button to move the microplate to the currently saved second side position
and switch to "teach-in" mode. See
Teach-In the Second Side for Crosswise
Press this button to increase the offset value by 0.1 mm to a maximum of 5.0 mm.
Press this button to decrease the offset value by 0.1 mm to a minimum of –5.0 mm.
Press this button to accept the displayed offset value and then define the aspiration
height. See
Define the Height for Aspiration on page 105
.
Table 7-25: Actions of Control Panel Buttons
Summary of Contents for MultiWash+
Page 1: ...5032022 C August 2019 MultiWash Microplate Washer User Guide ...
Page 20: ...MicroWash Microplate Washer User Guide 20 5032022 C ...
Page 50: ...MicroWash Microplate Washer User Guide 50 5032022 C ...
Page 64: ...MicroWash Microplate Washer User Guide 64 5032022 C ...
Page 144: ...MicroWash Microplate Washer User Guide 144 5032022 C ...