Chapter 7: Defining a Microplate Type
5032022 C
101
Define the Side for Aspiration
This option defines the position of a flat-bottom microplate for aspiration. If crosswise
aspiration is enabled, then this option defines the position for the first step of a crosswise
aspiration. Molecular Devices recommends that you place the aspiration needle close to the
edge the well in a flat-bottom microplate.
Figure 7-29: Aspiration position
You can define the side for aspiration by editing the offset value, or by using the "teach-in"
method.
Figure 7-30: The Side Position Front prompt
Button
Action
On the left
Press this button to end the process and return to
Define Plate
without saving any
changes. See
Defining a Microplate Type on page 77
.
On the right
Press this button to move the microplate to the currently saved aspiration position
and switch to "teach-in" mode. See
Teach-In the Side for Aspiration on page 102
.
Press this button to increase the offset value by 0.1 mm to a maximum of 5.0 mm.
Press this button to decrease the offset value by 0.1 mm to a minimum of –5.0 mm.
Press this button to accept the displayed offset value.
If crosswise aspiration is enabled, then you can define the second side for aspiration.
See
Define the Second Side for Crosswise Aspiration on page 103
.
If crosswise aspiration is disabled, then you can define the aspiration height. See
Define the Height for Aspiration on page 105
Table 7-23: Actions of Control Panel Buttons
Summary of Contents for MultiWash+
Page 1: ...5032022 C August 2019 MultiWash Microplate Washer User Guide ...
Page 20: ...MicroWash Microplate Washer User Guide 20 5032022 C ...
Page 50: ...MicroWash Microplate Washer User Guide 50 5032022 C ...
Page 64: ...MicroWash Microplate Washer User Guide 64 5032022 C ...
Page 144: ...MicroWash Microplate Washer User Guide 144 5032022 C ...