User Guide
EN
Dansensor
®
MAP Check 3
Vacuum
General setup
Selecting
General setup
from the
Main menu
will display a menu with available setup
parameters.
“Supervisor” access level is required for this menu, as the parameters in this menu control the
basic functionality of the device.
The
General setup
menu holds the following items:
Autostart
No
Device must be started manually using the
“Measure” button on the display
Yes
The device starts automatically when heating
period ends.
Disable faults
No
All faults are detected and READY relay contacts
will be activated to stop packaging machine logic.
Yes
All faults are detected but external signalling is
disabled.
This is used as last resort if you want to be able to
run packaging machine even though an error
exists in the device.
This setting will be indicated in the measure
screen by a warning symbol and a red text saying
"Faults disabled".
External mixer settings
Opens the
External mixer settings
screen
See
"External mixer settings" on page 61
for details.
Buffer measuring setup
Opens the
Buffer measure setup
screen
See
"Buffer measure setup" on page 62
for details.
Cycle counter
Can be set to any number. Counter is set to 0 from the factory.
Can be set to match the packaging machine cycle counter.
Data log setup
Opens the
Data log setup
screen
See
for details.
Network setup
Opens the
Network setup
screen
See
for details.
User label:
This item is a user configurable text that will be combined with
data logging information. Examples of use could be
manufacturing line identification or other similar data.
See
for details.
General setup
Summary of Contents for AMETEK Dansensor MAP Check 3 Vacuum
Page 1: ...P N 320598 G 11 2019 UserGuide 7BDVVN...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 20: ...18 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 54: ...52 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 76: ...74 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 84: ...82 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 91: ...This blank page has been inserted to enable double sided printing of the document...