Dansensor
®
MAP Check 3
Vacuum
EN
User Guide
43
92
Error
Ir CO2 sensor temperature too low, shutting down
CO2 sensor.
Check ambient
temperature -
must be within
range
If problem
persists, contact
service
93*
Warning
Ir CO2 sensor temperature high
Check ambient
temperature -
must be within
range
Check cooling/
filters
If problem
persists, contact
service
94
Error
Ir CO2 sensor temperature too high, shutting down
CO2 sensor.
Check ambient
temperature -
must be within
range
Check cooling/
filters
If problem
persists, contact
service
98
Error
Input pressure too high
Reduce pressure
at gas input
99
Error
Input pressure too low
Increase pressure
at gas input
101
Warning
Product uses external mixer but mixer is disabled
Enable mixer or
choose different
product
102
Warning
Buffer alarm
Check alarm
settings for
buffer.
103
Event
TIMING PROBLEMS, VAC signal too soon
Check control
signals
104
Event
TIMING PROBLEMS, IR delay too long
Check control
signals
105
Error
Critical high pressure on buffer input. More than 10
bar may be destructive
Lower input
pressure on
buffer input
232
Error
Vacuum is above max. Vacuum=%s, Max=%s
Check vacuum
No.
Type
Message
User action
Summary of Contents for AMETEK Dansensor MAP Check 3 Vacuum
Page 1: ...P N 320598 G 11 2019 UserGuide 7BDVVN...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 20: ...18 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 54: ...52 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 76: ...74 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 84: ...82 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 91: ...This blank page has been inserted to enable double sided printing of the document...