Dansensor
®
MAP Check 3
Vacuum
EN
User Guide
25
NOTE! If the device is installed on a packaging machine without vacuum, the
“EVAC” and “GAS” signals must be connected together to the gassing signal.
NOTE! All “I/O CONTROL” signals are galvanic isolated.
5 - Grey
6 - Pink
"ALARM", NO
Alarm relay output. Activated if product or buffer alarm limit
for O
2
/CO
2
is exceeded.
Relay contacts: max. 48 VDC/VAC. Load: max. 1A
Normally Open (NO) - Contacts closed during power OFF.
5 - Grey
7 - Blue
"ALARM", NC
Alarm relay output. Activated if product or buffer alarm limit
for O
2
/CO
2
is exceeded.
Relay contacts: max. 48 VDC/VAC. Load: max. 1A
Normally Closed (NC) - Contacts open during power OFF.
8 - Red
9 - Black
"WARNING", NO
Warning relay output. Activated if product or buffer alarm limit
for O
2
/CO
2
is exceeded.
Relay contacts: max. 48 VDC/VAC. Load: max. 1A
Normally Open (NO) - Contacts closed during power OFF.
8 - Red
10 - Purple
"WARNING", NC
Warning relay output. Activated if product or buffer alarm limit
for O
2
/CO
2
is exceeded.
Relay contacts: max. 48 VDC/VAC. Load: max. 1A
Normally Closed (NC) - Contacts open during power OFF.
11 - Grey/Pink
12 - Blue/red
"READY", NO
Ready relay output. Activated when device is READY (heating
finished, vacuum OK and device started).
Relay contacts: max. 48 VDC/VAC. Load: max. 1A
Normally Open (NO). Contacts closed during power OFF.
11 - Grey/Pink
13 - White/Green
"READY", NC
Ready relay output. Activated when device is READY (heating
finished, vacuum OK and device started).
Relay contacts: max. 48 VDC/VAC. Load: max. 1A
Normally Closed (NC) - Contacts open during power OFF.
Pins/Colours/
Function
Description
Summary of Contents for AMETEK Dansensor MAP Check 3 Vacuum
Page 1: ...P N 320598 G 11 2019 UserGuide 7BDVVN...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 20: ...18 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 54: ...52 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 76: ...74 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 84: ...82 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 91: ...This blank page has been inserted to enable double sided printing of the document...