![Mocon AMETEK Dansensor MAP Check 3 Vacuum User Manual Download Page 50](http://html1.mh-extra.com/html/mocon/ametek-dansensor-map-check-3-vacuum/ametek-dansensor-map-check-3-vacuum_user-manual_1816849050.webp)
48
User Guide
EN
Dansensor
®
MAP Check 3
Vacuum
Test modes in Diagnostics
Test measuring - Cal inlet
This test is used for calibration of the sensors in the system. It is normally used for servicing
purposes.
Remove the blind plug from the input labelled “CAL. GAS IN” and connect a suitable M5
fitting.
Connect a calibration gas to the input.
The gas must be regulated using a needle valve for a flow of 0.5 L/min ± 0.1 L/min.
Apply vacuum to the device and stop the packaging machine.
Enter
Diagnostics
and go to page 2 and select
Test measuring - Cal inlet
in the menu.
The device starts a continuous measurement of the connected calibration gas.
It is now possible to close the
Diagnostics
window and to go to the measuring screen to
watch the gas values.
The measuring screen will say “Test measuring”.
The test will run until you press
Stop
or until
Test measuring - Cal inlet
is pressed again on
page 2 in
Diagnostics
.
NOTE! Always install the blind plug in the “CAL. GAS IN” input when not in use.
25: Meat Mix 1000gr.
Cycles
6250
Test measuring
O
2
Bal
CO
2
%
%
%
Buffer
09:42
%O
2
%CO
2
bar
21,0
0,0
2,2
Summary of Contents for AMETEK Dansensor MAP Check 3 Vacuum
Page 1: ...P N 320598 G 11 2019 UserGuide 7BDVVN...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 20: ...18 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 54: ...52 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 76: ...74 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 84: ...82 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 91: ...This blank page has been inserted to enable double sided printing of the document...