![Mocon AMETEK Dansensor MAP Check 3 Vacuum User Manual Download Page 43](http://html1.mh-extra.com/html/mocon/ametek-dansensor-map-check-3-vacuum/ametek-dansensor-map-check-3-vacuum_user-manual_1816849043.webp)
Dansensor
®
MAP Check 3
Vacuum
EN
User Guide
41
52
Error
O
2
sensor is disabled because of hardware fail.
Please call service.
Contact service
54
Warning
Error in O
2
sensor calibration data. Check if the gas
concentration and span is correct.
Contact service
for sensor
calibration
55
Warning
O
2
sensor needs calibration. Measuring values may
not be valid.
Contact service
for sensor
calibration
58
Error
Fan speed too low, check if fan is functional
Contact service
59
Error
Error in fan parameters, running fan at full speed.
Contact service
60
Error
Temperature sensor not calibrated. Device
temperature may not be valid.
Check cooling/
filters
If error persists,
contact service
69
Error
Sample system clogged. Check hose for blockages
User information
70
Error
Illegal mixer setting: a 2-gas mixer is connected but
3-gas mixer is expected
User information
71
Error
Cannot establish connection to external mixer
User information
72
Error
Data log network dump failed: cannot connect to
server
Check LAN
connection
73
Error
Data log network dump failed: connection to server
was lost
Check LAN
connection
74
Error
Data log network dump failed: no ACK received
Check LAN
connection
75
Error
Data log network dump failed: wrong ACK received Check LAN
connection
76
Warning
Mixer low pressure alarm on gas input
X
User information
77
Error
Error from slave mixer
User information
79
Error
Zr O
2
cold-junction temperature too high. Possible
O
2
sensor fan fault.
Contact service
83
Warning
Ambient pressure value out of reasonable range.
Possible pressure sensor fault.
Check and clean
Air inlet/Sensor
gas/sample gas
outlets
If error persists,
contact service
No.
Type
Message
User action
Summary of Contents for AMETEK Dansensor MAP Check 3 Vacuum
Page 1: ...P N 320598 G 11 2019 UserGuide 7BDVVN...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 20: ...18 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 54: ...52 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 76: ...74 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 84: ...82 User Guide EN Dansensor MAP Check 3 Vacuum P N 320598 G 11 2019...
Page 91: ...This blank page has been inserted to enable double sided printing of the document...