9 Troubleshooting
Error
Possible cause
Way of elimination
The Stainer does not
switch on.
1 The fuses have blown.
1 Change the fuses. The fuses can be
changed only when the Stainer is switched
off (plug of the power cord must be
disconnected from the socket).
Remove lid from the fuses to change them
(12) (
Fig. 3).
2 The power cord is faulty.
2 Change the power cord.
Error when the
manipulator grabs a
rack.
Disruption of geometry of
either a rack or a parking
station (trough).
Detect the faulty accessory with the help of
the
Positioning test.
Change the faulty
accessory.
Error during the loading
of racks.
1 Incorrect placement of
troughs (positioning
elements do not fit into the
notches of the positioning
disk).
1 Install the parking stations correctly.
2 Lids are not removed from
troughs.
2 Remove the lids from the troughs.
Overflow of the flow-
through trough,
actuation of the water
level sensor.
The Stainer emits an
intermittent sound alarm
with a message on the
screen: <WATER
SYSTEM IS
CORRUPTED>.
1Incorrect laying of the drain
hose.
Check how the drain hose is lain (it must
be lain and fixed in such a way that all its
sections would be lower than the exhaust
water outlet).When laying the hose, make
shure there are no revers slopes of hose.
2 Obstruction of the drain
Clean or change the hose (
Item 8.2
).
Any problems and queries should be referred to your supplier. In other cases, refer
to the manufacturer company for repair of the Stainer.
CAUTION!
BEFORE SENDING THE STAINER FOR REPAIR, IT MUST BE
APPROPRIATELY CLEANED AND DISINFECTED (
.
Chapter 11
).
10 Decommissioning and recycling
CAUTION!
BEFORE DECOMMISSIONING AND RECYCLING, THE STAINER
MUST BE CLEANED AND DISINFECTED (
.
Chapter 11
).
CAUTION!
THE STAINER AND ITS PARTS MUST BE RECYCLED UNDER
REQUIREMENTS, EXISTING IN THE COUNTRY, WHERE THE RECYCLING TAKES
PLACE.