Figure 22. Configuration of station No. 3 with reagent.
- <STATION APPLICATION> parameter <TECHNOLOGICAL> – editable parameter.
The station is reserved for a trough with reagent. On pressing the button
TECHNOLOGICAL, the value is switched to <PARKING>, which means that the
station is now designated for placement of racks. The only editable parameter of
parking stations is <STATION APPLICATION>.
- <REAGENT> parameter <BUFFER PH<7> (name of reagent) – editable parameter.
If necessary, the name of a reagent can be changed by pressing the button with the
parameter’s name and selecting a needed name from the list. To enter a new
reagent, press the button NEW (reagents screen No. 5). The manufacturer’s list of
reagents is given in
Appendix B
.
- <RESOURCE LIMIT> – editable parameter: how many times it is left for the
manipulator to use a station with reagent until the exhaustion of its limit (e.g.,
because of changing properties of reagent). The admissible limit of usages is set by
the operator within the interval 0-255, default value is 255.
CAUTION!
WHEN THE VALUE REACHES ZERO, A MESSAGE APPEARS ON
THE SCREEN, WARNING ABOUT THE EXHAUSTION OF RESOURCE LIMIT OF
A REAGENT:
<RESOURCE LIMIT OF TROUGH # XX IS EXHAUSTED>
It gives the opportunity to duly change reagent in the respective trough. To continue
the process, press the button CANCEL. The value is reset after each reloading of
the Stainer.