Chapter Two: Installation
Setup
16
Mounting Screw Length
Note
The maximum screw penetration into the valve mounting flange is equal to the nominal screw
diameter. The use of screws exceeding this depth may result in improper flange preload and/or
damage to the valve.
Table 9: Mounting Screw Length
Nominal Size
Screw Quantity X Diameter
mm inch
ISO-F
JIS
320
250
200
160
100
12
10
8
6
4
NA
12 X M10
12 X M10
8 X M10
8 X M8
12 X M12
12 X M12
8 X M12
8 X M10
NA
Installation Procedure
1.
Install the control valve on the vacuum chamber. Chamber mounting side is indicated on the outline
drawing and is the side opposite the motor.
2.
Connect compressed air: Connections are for 1/4” tube
Use only clean, dry air or N2
Compressed air pressure (refer to Table 10 below)
Connect compressed air to controller connection “AIR/N2 INPUT”
3.
Connect the heating device (see
Appendix A: Product Specifications
4.
Option: Connect vacuum line to intermediate pumping port (1/8” NPT) rotary shaft seals.
Table 10: Required Air Pressure Based on Valve Size
Valve Size
Compressed Air Pressure
mm inch
psig
320
250
200
160
100
12
10
8
6
4
75-100
65-100
55-100
55-100
55-100
Summary of Contents for T3PIA
Page 3: ...MKS Type T3PIA Pendulum Valve With Analog TTL Interface...
Page 8: ...viii This page intentionally left blank...
Page 27: ...Electrical Information Chapter Two Installation 19 Figure 4 Top Panel of the T3P Unit...
Page 32: ...Chapter Two Installation Startup 24 This page intentionally left blank...
Page 60: ...Appendix C Dimensions 52...
Page 61: ...Appendix C Dimensions 53...
Page 62: ...Appendix C Dimensions 54...
Page 63: ...Appendix C Dimensions 55...
Page 64: ...Appendix C Dimensions 56...
Page 65: ...Appendix C Dimensions 57...
Page 66: ...Appendix C Dimensions 58...
Page 67: ...Appendix C Dimensions 59...
Page 68: ...Appendix C Dimensions 60...
Page 70: ...Index 62...