55
Signum 3 Ex
Averaging (Animal Weighing) Application
V
Averaging (Animal Weighing) Application
V
Preparation
§
Turn scale on: Press the
e
key
§
While all segments are lit, press the
)
key
§
Select Setup: Press the
k
key
repeatedly until
appl
is displayed
§
Confirm
appl
: Press the
)
key
§
Select the Animal Weighing application:
Press the
k
key repeatedly and
confirm with the
)
key
Application Parameters: Animal Weighing
3� 6� Minimum load for start
3�6�1* 1
digit
3�6�2 2
digits
3�6�3 5
digits
3�6�4 10
digits
3�6�5 20
digits
3�6�6 50
digits
3�6�7 100
digits
3�6�8
200 digits
3�6�9 500
digits
3�6�10 1000
digits
3�18� Start of averaging
3�18�1* Manually
3�18�2 Automatic
3�19� Animal activity
3�19�1
0�1% of animal/object
3�19�2* 0�2% of animal/object
3�19�3
0�5% of animal/object
3�19�4
1% of animal/object
3�19�5
2% of animal/object
3�19�6
5% of animal/object
3�19�7
10% of animal/object
3�19�8
20% of animal/object
3�19�9
50% of animal/object
3�19�10 100% of animal/object
3�20� Auto printout of results
3�20�1* Off
3�20�2 On
3�21� Static display of result after load
removed
3�21�1* Display is fixed until
unload threshold
reached
3�21�2
Fixed display until the
c
key is pressed
* = Factory setting
§
Save the settings with the
)
key and
exit Setup: Press the
(
key several
times�
Minimum Load
The minimum load required for
initialization of the averaging routine
is configured in Setup under:
appl1
:
anim.wg
:
3.6.
Setting a minimum load for averaging
can be especially useful if you configure
automatic start of measurement�
The minimum load required for
automatic taring of the container
weight on the platform (first weight),
or for automatic printout of results,
is configured in Setup under:
appl
:
a.tare
:
3.5.
You can choose from the following
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform� If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (= 1000 intervals)
on the weighing platform to start the
averaging routine�
Starting the Measurements
The averaging routine does not begin
until the fluctuation in weight value
remains below a defined threshold
over three consecutive measurements�
The tolerance limit is defined as a
percentage of the animal or object
weight (for example, 0�1%, 0�2%, ���,
50%, 100%), configured in Setup under:
appl1
:
anim.wg
:
3.19.
If the “Averaging” parameter is set to
2%, for example, and the animal or
object weighs 10 kg, measurement
does not begin until the fluctuation
in weight value remains below 200 g
during three consecutive measurements�
Display
A calculated average value is shown
continuously on the main display�
The
!
Symbol (indicating a calculated
value) is also displayed�
You can toggle between this display to
a readout of the current weight on the
platform by pressing the
w
key�
In the Setup menu, under:
appl1
:
anim.wg
:
3.21.
you can select “Display is static until
unload threshold reached” to have the
display switch automatically to the
weight readout when you unload the
weighing platform (i�e�, when the load
is less than half the minimum load)�
The result of the most recent averaging
operation is not saved�
If you select “Display is static until
the
c
key is pressed,” the calculated
average remains displayed even after
the weighing platform is unloaded,
until you press the
c
key to begin
a new measurement�
Summary of Contents for Signum 3 Ex
Page 97: ...97 Signum 3 Ex 97 Declaration of Conformity...
Page 98: ...98 Signum 3 Ex Declaration of Conformity...
Page 99: ...99 Signum 3 Ex 99 Certificates 2017 11 23...
Page 100: ...100 Signum 3 Ex Certificates...
Page 101: ...101 Signum 3 Ex Certificates...
Page 102: ...102 Signum 3 Ex Certificates...
Page 104: ...104 Signum 3 Ex Certificates...
Page 113: ...113 Signum 3 Ex...
Page 117: ...117 Signum 3 Ex Verification of intrinsic safety...
Page 118: ...118 Signum 3 Ex Verification of intrinsic safety...
Page 119: ...119 Signum 3 Ex Verification of intrinsic safety...
Page 120: ...120 Signum 3 Ex Verification of intrinsic safety...
Page 121: ...121 Signum 3 Ex Verification of intrinsic safety...
Page 122: ...122 Signum 3 Ex Verification of intrinsic safety...
Page 123: ...123 Signum 3 Ex Verification of intrinsic safety...
Page 124: ...124 Signum 3 Ex 124...
Page 134: ...134 Signum 3 Ex...
Page 135: ...135 Signum 3 Ex...