60
Signum 3 Ex
Checkweighing Application
O
Checkweighing
Application
O
With the Checkweighing application,
you can check whether the sample on
the weighing platform matches a target
value or lies within a given tolerance
range�
Checkweighing also makes it easy to fill
sample materials to a specified target
weight�
Characteristics
– Enter the nominal or target weight (set
point) and the tolerance range delimiters
either using the keypad or by saving the
weight value of a load on the platform�
– Enter the tolerance limits as absolute
values (Target, Min and Max) or via the
target value as percentage deviation
with user-defined percent limits or
with relative limits�
Configured in the menu under:
Appl 2
:
check.wg
:
4.5.
– The target value can be taken over
as a weighed value from a weighing
platform, and the upper and lower
tolerance limits are defined as a
percentage deviation from the target
value (setting code 4�5�2)� The deviation
percentage can be changed using the
r
key: 0�1%, 0�2%, 0�5%, 1%, 1�5%,
2%, 3%, 5% or 10%�
– The target value, lower tolerance limit
(minimum) and upper tolerance limit
(maximum) can be applied as weighed
values from the weighing platform�
– Target value and tolerance limits
checked during input; values must
conform to: upper limit
>
target
>
lower limit
>
1 digit�
– Checkweighing range: either 30% to
170% of the target, or from 10% to
infinity�
– Results are shown on main display on
the bar graph, as well as sent to control
output ports for further processing�
– Toggle the main display between weight
and tolerances limits by pressing the
w
key� If the weight in the readout is
outside the tolerance range, “
LL
” (too
low) or “
HH
” (too high) is displayed�
– Activate Info mode with the
I
key
– Automatic printout configured in Setup
under:
Appl 2
:
check.wg
:
4.6.
– Automatic taring of container weight�
Configured in Setup under:
Appl
:
a.tare
:
3.7.
– Automatic initialization when you
switch on the scale with most recently
saved application data� Configured in
Setup under:
appl
:
a.start
:
3.8.
– You can assign different functions to
the
c
key for deleting applications�
When you clear applications, you can
delete either the data stored for all
applications or just selected data stored
for the active application� Configured in
Setup under:
Appl
:
sel.cf
:
3.24.
(Sel� CF function key
c
in
applications)
– Tare function:
1) If you store a tare (weight value)
by pressing the
)
key, you can later
enter a tare value manually� The tare
value you enter is added to the stored
tare value� Setting: menu code 3�25�1
(factory setting)
2) A tare value entered manually
overwrites a stored tare value
(weight value)�
If you enter a tare value manually,
a tare value (weight value) stored later
overwrites the manually entered value�
Setting: menu code 3�25�2
Configured in Setup under:
Appl
:
tare. f
:
3.25.
– Restore factory default settings�
Configured in Setup under:
Appl
:
def.app
:
9.1.
Checkweighing entails comparing the
current weight value to a defined target
value� You can enter the value for this
target using the keypad, or by saving
the weight value indicated� You can
also define upper and lower tolerance
limits based on this target� You can do
this by:
– Entering absolute values using the
keypad or placing the desired amount
of weight on the platform and saving
the value, or
– by entering each value as a percentage of
the target weight
The initialization data remains valid until
deleted by pressing the
c
key or until
overwritten by a new value� They remain
saved after the scale is switched off�
Preparation
§
Turn scale on: Press the
e
key
§
While all segments are lit, press the
)
key
§
Select Setup: Press the
k
key repeatedly
until
Setup
is displayed
§
Open the Setup menu: Press the
)
key
§
Select the Checkweighing application:
Press the
k
key repeatedly and confirm
with the
)
key
Application parameters: Checkweighing
4�2� Checkweighing range
4�2�1* 30 to 170%
4�2�2
10% to infinity
4�3� Control output for SET as:
4�3�1* Output
“SET”
4�3�2
Ready to operate
4�4� Activation of outputs
4�4�1 Off
4�4�2 Always
on
4�4�3
On at stability
4�4�4* On within checkweighing range
4�4�5
On at stability within
checkweighing range
4�5� Parameter input
4�5�1
Min, max, target value,
absolute values
4�5�2* Only target value with percent
limits
4�5�3 Target value with free percent
limits
4�5�4 Target value; min, max values
as relative weights to target
value
4�6� Automatic
printing
4�6�1* Off
4�6�2 On
4�6�3
Only values within tolerance
4�6�4
Only values outside tolerance
4�7� Checkweighing toward zero
4� 7� 1* Off
4�7�
2 On
* = Factory setting
§
Save the settings with the
)
key and exit
Setup: Press the
(
key several times�
Application 2: Checkweighing
O
Summary of Contents for Signum 3 Ex
Page 97: ...97 Signum 3 Ex 97 Declaration of Conformity...
Page 98: ...98 Signum 3 Ex Declaration of Conformity...
Page 99: ...99 Signum 3 Ex 99 Certificates 2017 11 23...
Page 100: ...100 Signum 3 Ex Certificates...
Page 101: ...101 Signum 3 Ex Certificates...
Page 102: ...102 Signum 3 Ex Certificates...
Page 104: ...104 Signum 3 Ex Certificates...
Page 113: ...113 Signum 3 Ex...
Page 117: ...117 Signum 3 Ex Verification of intrinsic safety...
Page 118: ...118 Signum 3 Ex Verification of intrinsic safety...
Page 119: ...119 Signum 3 Ex Verification of intrinsic safety...
Page 120: ...120 Signum 3 Ex Verification of intrinsic safety...
Page 121: ...121 Signum 3 Ex Verification of intrinsic safety...
Page 122: ...122 Signum 3 Ex Verification of intrinsic safety...
Page 123: ...123 Signum 3 Ex Verification of intrinsic safety...
Page 124: ...124 Signum 3 Ex 124...
Page 134: ...134 Signum 3 Ex...
Page 135: ...135 Signum 3 Ex...