7.5.3
Parameter P23: SetDeadload with weight (Write only)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
Byte 3
1
When writing into this parameter, the current weight will be saved as dead load.
7.5.4
Parameter P24: SetSpan (Write)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Weight MSB
Byte 1
Weight MSB
Byte 2
Weight LSB
Byte 3
Exponent
Unit
Calibration using weights
Weight
3 bytes for the weight value
Exponent
Number of decimal places
Unit
Weight unit (as in function number 4)
7.5.5
Parameter P25: Set/GetDeadloadMvpv (Write/Read)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Dead load MSB
Byte 1
Dead load MSB
Byte 2
Dead load MSB
Byte 3
Dead load LSB
DEAD LOAD in mV/V (value in mV/V) x 10
6
7.5.6
Parameter P26: Set/GetSpanMvpv (Write/Read)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
SPAN MSB
Byte 1
SPAN MSB
Byte 2
SPAN MSB
Byte 3
SPAN LSB
SPAN in mV/V (value in mV/V) x 10
6
Transmitter Series PR 5211
7 ProfiBus interface
Minebea Intec
EN-80