4.4.7.8
Connection to relay PR 1626/6x
The connection is made via the connecting cable PR 6135/.. .
The internal load cell power supply (V+, V-) of the PR 5211 must
not
be connected.
If supply voltage of <8 V DC is required for the load cells, the following settings must be
done:
Device
Component
Action
PR 5211/xx
Solder jumper J300 (see
Figure)
must be open on the A/D converter Print
(remove 0
Ω
impedance).
Note:
In the factory setting, the solder jumper is closed by a 0
Ω
impedance.
By opening J300 the sense voltage monitoring will be switched off!
NOTICE
Installation in the Ex zone
The screen of the load cell cable and the screen of the connecting cable must not be
connected inside the junction box, if connection of both ends is not permissible
according to the regulations for installation in the ex-zone.
4 Device installation
Transmitter Series PR 5211
EN-31
Minebea Intec