7.3.1
Configuration parameters
7.3.1.1
Parameter 1: Analog output
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
Range
Output mode
Byte 3
ADU in errorstate
ADU below zero
ADU above FSD
FSD = Full scale deflection
Analog output
Description
Range/Signal
Factory setting
Output mode
off,
(func. 30)*,
Gross,
Net
0,
1,
8,
9
0
Analog range
0…20 mA
4…20 mA
0
1
1
ADU in
error state**
hold
0 mA
4 mA
20 mA
0
1
2
3
1
ADU below zero
Linear ***
0 mA
4 mA
20 mA
0
1
2
3
0
ADU above SKE
Linear***
0 mA
4 mA
20 mA
0
1
2
3
3
* Output value 1 means that the value in function number 30 is written to the analog
output.
** The ADU error state is only valid if output value is set to gross (8) or net (9).
*** The linear selection can be used to get a proportional output signal outside the range
0 … FSD.
7.3.1.2
Parameter 2: Scaling weight value for 0/4 mA output
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
WEIGHT MSB
Byte 1
WEIGHT MSB
Byte 2
WEIGHT MSB
Byte 3
WEIGHT LSB
Transmitter Series PR 5211
7 ProfiBus interface
Minebea Intec
EN-74
Summary of Contents for PR 5211 Series