5.1.3.2
Removing the cover
Remove the cover as follows:
5.1.3.3
Factory settings
Calibration data <default>
Calibration parameters <default>
Full Scale (FSD) (Max) <3000> <kg>
Measurement time (M) <320> ms
Scale interval <1>
Measuring rate <160> ms
Dead load <0,000000> mV/V
Standstill time <1> M
SPAN <1,000000> mV/V
Standstill range <1,00> d
Standstill timeout <8> M
Calibration parameters <default>
Test mode <absolute>
Overload (range over Max.) <9>d
Zero range <50.00> d
W&M node <off>
Zerotrack range <0,25> d
Digital filter <off>
Zerotrack step width <0,25> d
Frequency <1,56 Hz>
Zerotrack time interval <0> M
In the original condition as delivered from the factory, the calibration data and
parameters are set to default values (Factory setting).
If a new calibration is started, the calibration data will be set to default (the parameters
remain unchanged).
5.2
Switching on the device
The device can be set up as follows:
-
Via a notebook/PC with the supplied program "ConfigureIt! 6.00" (or higher).
-
The ProfiBus interface allows access to individual parameters.
The weight display is shown.
1.
Insert a knife (1) into the slot below the cover and press until it is released.
2.
Remove the cap.
Transmitter Series PR 5211
5 Getting started
Minebea Intec
EN-40