4.4.7.6
Testing the measuring circuit
A simple test with the load cells connected can be carried out with a multimeter.
Note:
In the case of an external load cell supply voltage or use of an isolating unit, the internal
load cell supply is not relevant.
Supply voltage
12 V ±0.8 V (symmetrical to housing ground)
Sense voltage
12 V ±0.8 V (symmetrical to housing ground)
Measuring voltage
0–12 mV @ LC with 1.0 mV/V
0–24 mV @ LC with 2.0 mV/V
4 Device installation
Transmitter Series PR 5211
EN-29
Minebea Intec