Byte
Field
Description
5
Write
Active
Power
Fail
free
free
free
free
free
free
6
Cmd
Busy
Cmd
Error
free
free
free
Tare
Active
Cal
Chan-
ged
Test
Active
7
OutOf
Range
Stand-
still
Inside
ZSR
Center
Zero
Below
Zero
Over-
load
Above
Max
ADU
Error
In direct access, status bits are
independent of the write or read
request.
"Free" bits are application speci-
fic.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Field
Size
Function
Read data
4 bytes
Data to be read as a binary 32-bit value with plus or
minus sign.
Data type: DINT
Read_Value_Selected
1 byte
Acknowledgment of the transmitted function number.
WriteActive
1 bit
The function selected with
Write_Value_Select
has
been executed once.
This bit is deleted if
Write_Value_Select
is set to 0.
PowerFail
1 bit
Is set when switching on the device.
Is reset by
ResPower
with transition from 0→1.
CmdBusy
1 bit
The device is busy executing a function (e.g., waiting
for a standstill for taring)
CmdError
1 bit
The device has interrupted the execution of a com-
mand (e.g.,
standstill
could not be reached within the
defined standstill time).
The error number can be read from "LASTERROR", see
Chapter
Tare_Active
1 bit
The scale has been tared.
Cal_Changed
1 bit
The device has been calibrated.
When this bit is 1, the weighing point parameters (EX-
PO/UNIT/STEP+FSD) must be read again.
Set after "Power on" and reset after reading the FSD.
Test_Active
1 bit
The device executes the ADC test. The read weight
value is not the gross value, but the test value.
OutOfRange
1 bit
Below zero or above max. (FSD).
Standstill
1 bit
The scale is stable.
InsideZSR
1 bit
The gross weight value is within the zero setting range.
CenterZero
1 bit
The weight value is within
center zero
(0 ±0.25 d).
BelowZero
1 bit
The weight value is negative (gross < 0 d).
Overload
1 bit
The weight value has exceeded the measuring range.
No valid weight data is specified (gross > FSD+over-
load).
7 ProfiBus interface
Transmitter Series PR 5211
EN-63