9. Fieldbus interface | 9.2 - Scale protocol | 112
Field
Size
Function
CenterZero
1 bit
The weight value is within
center zero
(0
±0.25 d).
BelowZero
1 bit
The weight value is negative (gross < 0 d).
Overload
1 bit
The weight value has exceeded the measur
ing range. No valid weight data is specified
(gross > FSD+overload).
AboveMax
1 bit
The weight value has exceeded Max (FSD),
but is still within Max + permissible overload
(gross ≤ FSD+overload).
ADCError
1 bit
AD conversion error, see Chapter
9.2.2 Reading and writing data with function numbers
9.2.2.1 Reading data
Procedure:
1. Write the function number as
Read_Value_Select
in byte 4 of the write
window (e.g., 9 = net weight).
2. Wait until
Read_Value_Selected
in byte 4 of the read window is equal to
Read_Value_Select
of the write window.
The requested value is available in bytes 0-3.
Action of the master
Slave reaction
Write function number to
Read_Value_S
elect
.
Write requested data in
Read_Data
(bytes
0-3).
Copy
Read_Value_Select
to
Read_Value_
Selected
.
Wait until
Read_Value_Selected
=
Read_
Value_Select
.
Read requested data in
Read_Data
(bytes
0-3).