6. Operating | 6.14 - Calibrating internal weighing point | 99
6.14.2 Performing a new adjustment
Requirements:
–
Overwrite protection is deactivated, see Chapter
: [Adjustment]
1. Click [New] in the task bar.
The data is first set to factory setting (default) and then the
adjustment is started.
2. Setting max. capacity [Max. capacity], see Chapter
3. Determining the scale interval [Scale interval], see Chapter
4. Determining the dead load [Dead load], see Chapter
.
5. Adjustment with weight [Adjusted at], see Chapter
.
or
6. Adjustment with mV/V value [Adjusted at], see Chapter
or
7. Adjustment with load cell data (SMART calibration) [Adjusted at], see
Adjusting with load cell data (SMART calibration)
.
8. Activate the overwrite protection after finishing the adjustment, see
.