ZS3 Instructions for Use
49
|
P a g e
Lock for
Up/Down
Lock - Red
Unlock
Figure 4-7: P8-3m Adult and Pediatric TEE Lock Operation
The TEE probes include sensors to monitor the temperature of the tip during use. While
imaging the following on-screen indicator is shown:
The angle of rotation shows the scan plane rotation while the temperature indicator shows the
temperature at the transducer face. This indicator can optionally be turned off if the
temperature is below 40.5ºC. Above that temperature the indicator will always be displayed. In
addition:
•
If the temperature reaches 41.5ºC the system will automatically freeze and display a warning
message. This warning message can be dismissed and imaging can continue. While operating
at this temperature is safe, close monitoring of the temperature is recommended. The
temperature indicator will show in yellow to indicate this.
•
If the temperature reaches 42.7ºC the system will automatically freeze imaging. The
transducer must be disconnected and allowed to cool off before additional imaging is
supported.