ZS3 Instructions for Use
15
|
P a g e
Figure 2.2: Control Panel
Powering On/Off
1.
Press and release the
Power
button (located above the transducer connector) to
turn the system on.
2.
Check the ZS3 display to ensure the startup screen is shown. When
initialization is complete, the system is ready for imaging.
3.
To Power Off press and release the
Power
button.
NOTE: When servicing the ZS3 system, always be sure to turn the circuit breaker to
the Off position.
Basic Measurements (Caliper Button)
1.
Press
Caliper
while image is frozen.
2.
Select desired measurement tool on menu & press a
Set
Key.
3.
Position first caliper with
Trackball
& press a
Set
Key.
Position second caliper with
Trackball
& press a
Set
Key. (Note: Pressing the
Select
Key
will toggle between active calipers in the set or series for fine tuning)
Basic Calculations (Calc Button)
1.
Press
Calc
while image is frozen.
2.
Select desired calc on menu & press a
Set
Key.
3.
Position caliper with
Trackball
& press a
Set
Key. (Note: Pressing the
Select
Key will toggle between active calipers in the set or series for fine tuning).
4.
Position additional caliper(s) with
Trackball
& press a
Set
Key. Calc will display on screen.
Quick Start Guide
Summary of Contents for Zonare ZS3
Page 1: ...I ZS3 Ultrasound System Instructions for Use ...
Page 6: ......
Page 12: ......
Page 38: ......
Page 74: ......
Page 78: ......
Page 86: ......
Page 110: ......
Page 128: ......
Page 146: ......
Page 150: ......
Page 158: ......
Page 160: ......
Page 164: ...ZS3 Instructions for Use 147 P a g e ...
Page 165: ...P N 046 012347 01 3 0 ...