The system complies with the EMC standard IEC 60601-1-2: 2014.
Intended Environments:
Professional healthcare facility environment
.
WARNING:
1.
The use of unapproved accessories may diminish system
performance.
2.
Use of components, accessories, probes, and cables other than
those specified may result in increased emission or decreased
immunity of system.
3.
Operation of system, in the case that the patient physiological
signal is lower than the minimum amplitude or value specified
in the product specifications, results may not be obtained
(results can be obtained when the HR is in the range of 30-250
bmp or when the QRS wave amplitude is between 0.5-5 mV.)
4.
Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they
are operating normally.
5.
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and
result in improper operation.
6.
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used
no closer than 30 cm (12 inches) to any part of the system,
including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could
result.
Summary of Contents for Zonare ZS3
Page 1: ...I ZS3 Ultrasound System Instructions for Use ...
Page 6: ......
Page 12: ......
Page 38: ......
Page 74: ......
Page 78: ......
Page 86: ......
Page 110: ......
Page 128: ......
Page 146: ......
Page 150: ......
Page 158: ......
Page 160: ......
Page 164: ...ZS3 Instructions for Use 147 P a g e ...
Page 165: ...P N 046 012347 01 3 0 ...