ZS3 Instructions for Use
117
|
P a g e
5. Press the
Undo
key to exit transducer diagnostics.
Notes: Probe check is provided for reference only, not for confirming a diagnosis.
Each time a transducer is connected to the main system or activated, the users are
recommended to implement probe check of transducer performance.
System Care and Cleaning
Before cleaning, turn off the AC circuit breaker on the ZS3 system to remove all power from the
unit.
To clean the LCD display:
•
Clean the glass using a soft cotton cloth lightly moistened with a watery solution or a
mild commercial glass-cleaning product suited for coated glass surfaces.
•
Wipe dry with a clean, dry, soft, lint-free cloth.
•
Take care not to damage or scratch the glass or LCD panel. Do not apply pressure on the
glass or LCD panel. Do not apply or spray liquid directly to the glass, panel, or cabinet as
excess liquid can cause damage to internal electronics. Apply the liquid to the cleaning cloth.
To clean the External Case:
•
Do not use disinfectants (such as glutaraldehyde) or acetone to clean any surfaces on the ZS3
system or its accessories.
•
Wipe the ZS3 surfaces with a safe disinfectant solution such as Sani-Cloth Plus or 50%
isopropyl alcohol and follow the disinfectant label instructions for use.
•
Do not spill or spray liquid directly on the control panel, LCD display or transducer connector.
•
Using soap and water or a mild disinfectant, gently wipe the surfaces of the ZS3 system
with a moistened cloth.
•
After each use, remove and dispose of any used cover/sheath. Wipe off any excess gel
from the transducer and clean it properly and air dry the ZS3 system.
Summary of Contents for Zonare ZS3
Page 1: ...I ZS3 Ultrasound System Instructions for Use ...
Page 6: ......
Page 12: ......
Page 38: ......
Page 74: ......
Page 78: ......
Page 86: ......
Page 110: ......
Page 128: ......
Page 146: ......
Page 150: ......
Page 158: ......
Page 160: ......
Page 164: ...ZS3 Instructions for Use 147 P a g e ...
Page 165: ...P N 046 012347 01 3 0 ...