ZS3 Instructions for Use
64
|
P a g e
Table 1 2D image measurements
Parameter
Range
Error
Distance
Full Screen
Within ±3%
Area
Full Screen
Within ±7%
Circ
Full Screen
Within ±10%
Angle
Full Screen
Within ±3%
Volume
Full Screen
Within ±10 %
iScape (for linear,
and wide convex
probes)
Full Screen
Within ±5%
Table 2 3D/4D image measurements
Parameter
Range
Error
Distance
Section A/B/C
Within ±5% (not including
Smart 3D mode)
Area
Section A/B/C
Within ±7% (not including
Smart 3D mode)
Circ
Section A/B/C
Within ±10% (not including
Smart 3D mode)
Angle
Section A/B/C
Within ±5% (not including
Smart 3D mode)
Volume
Section A/B/C
Within ±20 % (not including
Smart 3D mode)
Table 3 1D image measurements
Parameter
Range
Error
Distance
Full Screen
Within ±3%
Time
Timeline Display
Within ±2%
Heart Rate
Timeline Display
Within ±4%
Velocity (PW)
10-200cm/s (for C4-1, C10-
3, C6-2, C9-3, C8-3 3D,
E9-3, E9-4, E9-3 3D, L10-
5, L8-3, L14-5sp, L14-5w,
L20-5, P8-3TEE, P9-3ic,
C9-3sp, C6-1, C18-5, P8-
3mTEE, L30-8, C12-4lp,
P4-1c, and P7-3c probes)
When angle
≤
60
º
,
≤
5%.
Measurement Accuracy
Summary of Contents for Zonare ZS3
Page 1: ...I ZS3 Ultrasound System Instructions for Use ...
Page 6: ......
Page 12: ......
Page 38: ......
Page 74: ......
Page 78: ......
Page 86: ......
Page 110: ......
Page 128: ......
Page 146: ......
Page 150: ......
Page 158: ......
Page 160: ......
Page 164: ...ZS3 Instructions for Use 147 P a g e ...
Page 165: ...P N 046 012347 01 3 0 ...