Monitor Description
1-1
Chapter 1
Monitor Description
1.1 Intended Use
This Monitor is intended for monitoring the patient’s vital signs including Non-invasive Blood
Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR) and Temperature (TEMP) for
single adult, pediatric and neonatal patient.
This Monitor is intended for use in the health-care institutions such as Outpatient Clinics,
Emergency Departments, Medical Floors, Clinics and Nursing Departments. It, however, is not
intended for critical patient monitoring, hospital transport or home use.
This Monitor is to be operated by clinical physicians or appropriate medical staffs under the
direction of physicians. The operator of the monitor must be well tranined. Any operation by
unauthorized or non-tranined personnel is forbidden.
1.2 Environmental Specifications
1.2.1 Temperature Requirements
Operating ambient temperature
0
ºC
- 40
ºC
10
ºC
~+40
ºC
(
50ºF -104ºF
)
(SmarTemp™ TEMP
module)
Storage temperature
-20
ºC
– 60
ºC
1.2.2 Relative Humidity
Operating relative humidity
15%-95%(non-condensing)
Storage relative humidity
10%-95%(non-condensing)
1.2.3 Altitude Requirements
atmospheric pressure at working conditon
(operating altitude)
70 to 106KPa
-500 to 4600m, -1640 to 15092ft
atmospheric pressure in Storage
(Storage altitude)
22 to 107.4KPa
-500 to 13100m, -1640 to 42979ft
Summary of Contents for VS-800
Page 1: ...VS 800 Vital signs monitor Service Manual ...
Page 2: ......
Page 10: ...Content II For your notes ...
Page 32: ...Principles of Operation 2 20 For your notes ...
Page 60: ...Machine Disassembly Assembly and Troubleshooting 4 20 For your notes ...
Page 68: ...Machine test and Material List 5 8 For your notes ...
Page 72: ...Maintenance and Cleaning 6 4 FOR YOUR NOTES ...
Page 73: ......
Page 74: ...P N 6006 20 39439 2 0 ...