Machine Disassembly/Assembly and Troubleshooting
4-17
Cannot start
Check if the AC indicator
is lit
Check the keypad-to-main board signal line for fault
;
Check the keypad-to-power board signal line for fault;
Check the keyboard for fault
Check the main board for fault;
Check the power board for fault;
Check/replace the
above units one by one
so as to locate the
fault;
Check if the power
supply is normal
Y
N
Check the AC power
cord is norma
l
Y
Check the keypad-to-power board signal
line for fault;
Check the keyboard for fault
Check the main board for fault;
Check the power board for fault;
Y
Check if battery is
installed;
Check if the battery
has electricity;
N
Y
Y
N
N
N
end
Figure 25
Black Screen Troubleshooting Map
4.3.2 LCD Black, Backlight Off
1. Check if the LCD signal line is in good condition;
2. Check if the keypad power cord is in good condition;
3. If the problem still remains after the connection cables are replaced, replace the
LCD module:
4. If the problem still exists, check the keypad, power board or main board.
4.3.3 LED Displays Abnormally
1.
Check if the LED signal line is in good condition;
2.
Check if the keypad power cord is in good condition;
3.
Check if the button signal line is in good condition;
4.
If the problem still exists, check the keypad, power board or main board.
4.3.4 Alarms Soundless
1.
Check if the sound is turned off in the software setup;
Summary of Contents for VS-800
Page 1: ...VS 800 Vital signs monitor Service Manual ...
Page 2: ......
Page 10: ...Content II For your notes ...
Page 32: ...Principles of Operation 2 20 For your notes ...
Page 60: ...Machine Disassembly Assembly and Troubleshooting 4 20 For your notes ...
Page 68: ...Machine test and Material List 5 8 For your notes ...
Page 72: ...Maintenance and Cleaning 6 4 FOR YOUR NOTES ...
Page 73: ......
Page 74: ...P N 6006 20 39439 2 0 ...