Product Specifications
3-6
Max. mean error
Max. standard error
±
5mmHg
8mmHg
Software over-pressure protection: The over-pressure detection is controlled by
software. Once the cuff pressure exceeds the threshold, the software enables the system
to deflate the cuff.
Adult 300
mmHg
Pediatric
243mmHg
Neonate 150
mmHg
Hardware over-pressure protection: In single fault conditions, the hardware
controls the cuff deflation to prevent the cuff pressure from exceeding the following ranges.
Adult 330
mmHg
Pediatric
330mmHg
Neonate 165
mmHg
3.2.9 SpO
2
3.2.9.1 MASIMO SpO
2
Range
SpO2 (%)
1% - 100%
PR (bmp)
25 - 240
Accuracy
SpO
2
(%)
No motion conditions
Adult/pediatric
70% - 100%: ±2%
0% - 69%: Undefined
Neonate
70% -100%: ±3%
0% - 69%: Undefined
SpO
2
(%)
During motion conditions
Adult/pediatric/neonate
70% -100%: ±3%
0% - 69%: Undefined
PR (bpm)
No motion conditions
25 - 240 ± 3 BPM
PR (bpm)
During motion conditions
25 - 240 ± 5 BPM