Product Specifications
3-5
3.2.8 NIBP
Measuring method
Automatic oscillometry
Mode of operation
Manual/Automatic/Continuous
Measuring interval for automatic
mode:
1,2,3,4,5,10,15,30,60,90,120,180,240,480min
Measuring time for continuous
mode:
5min
Maximum measurement cycle
Adult/pediatric:
Neonate:
180s
90s
PR:
Range
Accuracy
Resolution
40 - 240 bpm
±
2 bpm or
±
2%, whichever is the greater
1bpm
Measurement range in normal mode
Adult:
Systolic pressure:
40 - 270mmHg
Diastolic pressure:
10 - 210mmHg
Mean pressure:
20 - 230mmHg
Pediatric:
Systolic pressure:
40 - 200mmHg
Diastolic pressure:
10 - 150mmHg
Mean pressure:
20 - 165mmHg
Neonate :
Systolic pressure:
40 - 135mmHg
Diastolic pressure:
10 - 100mmHg
Mean pressure:
20 - 110mmHg
Resolution:
Pressure:
1mmHg
Accuracy:
Pressure