Maintenance and Cleaning
6-1
Chanpter6
Maintenance and Cleaning
6.1 Inspection
6.1.1 Inspection before Using the Monitor
Before using the monitor, check:
1.
The monitor for mechanical damages;
2.
All exposed wires, male connectors and accessories;
3.
All functions that may be used on patients, thus guaranteeing the good performance
of the monitor.
In case of any indication of functional damage, stop using the monitor, and contact bio-medical
engineers of the hospital or Mindray service engineers immediately.
6.1.2 Routine Inspection
An overall inspection, including the functional safety inspection, must be performed on the
monitor by qualified personnel for every 6-12 months or after maintenance each time. All
inspections must be performed by qualified service personnel when it is necessary to
disassemble the monitor.
Warning
Failure on the part of the responsible hospital or institution employing the
monitoring equipment to implement a satisfactory maintenance schedule may
cause undue equipment failure and possible health hazard.