Product Specifications
3-3
Maximum load current
≤
2A
Connection resistance
<1
Ω
Isolation voltage
>1500V AC
Network
interface
For connection to the Central Monitoring
Network, software upgrade and data export
3.2.5 Battery
No. of batteries:
1
Battery type:
Sealed lead-acid battery or lithium ion battery
Time to shutdown at low battery power: 5-15min after the first low-battery alarm (a new
fully-charged battery should be used)
3.2.5.1 2.3 Ah 12V Sealed Lead-Acid Battery
Minimum of 260min
The monitor runs on power supplied by the new fully-charged battery in the following
conditions:
Ambient temperature: 25
℃
Monitor configuration: SpO
2
(Continuous measurement) and NIBP (one auto NIBP
measurement per 15min)
Charge time: Maximum of 8h. Operating status.
3.2.5.2 4.4Ah 11.1V Lithium Ion Battery
Minimum of 620min
The monitor runs on power supplied by the new fully-charged battery in the following
conditions:
Ambient temperature: 25
℃
Monitor configuration: SpO
2
(Continuous measurement) and NIBP (one auto NIBP
measurement per 15min)
Charge time: Maximum of 8h. Operating status.
3.2.6 Recorder
Method
Thermal dot array
Horizontal resolution
160dots/cm (25 mm/s)
Vertical resolution
80dots/cm
No. of waveform channels
1
Paper width
50±0.1 mm
Summary of Contents for VS-800
Page 1: ...VS 800 Vital signs monitor Service Manual ...
Page 2: ......
Page 10: ...Content II For your notes ...
Page 32: ...Principles of Operation 2 20 For your notes ...
Page 60: ...Machine Disassembly Assembly and Troubleshooting 4 20 For your notes ...
Page 68: ...Machine test and Material List 5 8 For your notes ...
Page 72: ...Maintenance and Cleaning 6 4 FOR YOUR NOTES ...
Page 73: ......
Page 74: ...P N 6006 20 39439 2 0 ...