Product Specifications
3-4
Paper length
30m
Paper speed
25 mm/s, ±5%
Record mode
The recorder should support 3 record modes:
1) Recording current trend data being displayed on the LCD
of the current patient
2) Recording all patient trend data
Recording waveform graphs continuously in real time
3.2.7 Review
Trend review
Displayed contents of full configuration:
Data in trend tables of all working parameters, including systolic
pressure, mean pressure, diastolic pressure, pulse rate, SpO
2
,
measurement time and patient ID.
SpO
2
configuration only: Once per minute
(mean measured value within 1min.)
Display resolution:
Other configurations: Minimum interval is
1min.
Storage of trend data:
Maximum storage: 1200 groups of data
Storage of data
Storage of monitor configurations:
The monitor should be capable of storing changeable
configurations. Besides the factory default configuration, one
group of user default configuration for each patient type
should be stored.
The following data should be stored:
1. Patient
information
2. Alarm limit settings
3. Volume
settings
4. Recorder
settings
5. Clock
settings
6. Parameter
settings
Storage of patient information:
The current patient information is stored in the non-volatile
memory.
Summary of Contents for VS-800
Page 1: ...VS 800 Vital signs monitor Service Manual ...
Page 2: ......
Page 10: ...Content II For your notes ...
Page 32: ...Principles of Operation 2 20 For your notes ...
Page 60: ...Machine Disassembly Assembly and Troubleshooting 4 20 For your notes ...
Page 68: ...Machine test and Material List 5 8 For your notes ...
Page 72: ...Maintenance and Cleaning 6 4 FOR YOUR NOTES ...
Page 73: ......
Page 74: ...P N 6006 20 39439 2 0 ...