Figure 16 Save as DCM Format
2.2.3
Common Reasons of Failure and
Troubleshooting
Failure:
The server disconnects to the system automatically.
Potential Reasons:
The storage service of connected server cannot support JPEG compression, while which
is selected in storage service in the system.
Troubleshooting:
Modify the storage configuration of this server in Storage screen in the system. Select the
service and select “Uncompressed”, and then click [Update] to update the latest
configuration information.
Failure:
The net is disconnected during transmission.
Potential Reasons:
The connected net wire is plugged out during transmission.
Troubleshooting:
Reconnect net wire to ensure normal net connection status.
Failure:
Multi-frame transmission is failed, but single frame is succeeded.
Potential Reasons:
Enable multi-frame image storage is not selected in the dialog box of Storage preset.