mindray
DC7 Diagnostic Ultrasound System
36
Troubleshooting:
Set the right Magnification Type in local ultrasound DICOM Service preset according
to the Magnification Type specified in the printer conformance statement.
If there is no printer conformance statement for reference, set the Magnification Type
as Bilinear, CUBIC or REPLICATE.
3.8
Common Failure: MPPS message is
not sent out correctly or not received
by the server
Potential Reasons:
The MPPS server is not set as default.
Troubleshooting:
Set the MPPS server as default after setting the Port and AE Title correctly.
3.9
Common Failure: Storage
commitment is failed
Potential Reasons:
1. The associated storage commitment server is incorrect;
2. No default commitment server is set;
3. The port and IP address settings on the commitment server are inconsistent with
the local preset.
Troubleshooting:
1. Check if the associated storage commitment server is correct;
2. Check if a default commitment server is set;
3. Check if the port and IP address settings on the commitment server are
consistent with the local preset.
NOTE:
The storage commitment icon
in [iStation] screen is not refreshed in realtime. Exit and
enter the [iStation] screen will refresh the commitment icon if a storage commitment is
succeeded.
3.10 Common Failure: Query is succeeded,
but retrieve is failed, with the
prompting that the remote server is
wrong
Potential Reasons:
The setting of the local storage SCP port is wrong.
Troubleshooting: Some servers only support the specified port as the storage SCP
port. Please confirm whether the server has this restriction. If it has, please modify the
serial number of local port. See the Figure below: